A pipette calibration certificate can help confirm that an instrument meets the requirements for your work, but not every pipette needs accredited calibration. A micropipette can keep clicking, loading tips and dispensing while consistently delivering less than its set volume. By the time the error shows up in your results, you may have repeated experiments or blamed other equipment. Verification helps you find out whether the pipette is the source of the problem.

What a pipette calibration certificate tells you

Verification, calibration and adjustment are related, but they are not the same. Confusing them can lead to unnecessary service costs or a false sense that an instrument is covered.

  • Verification: checks whether the pipette dispenses within limits set by your laboratory. You can do this in-house using a balance and distilled water. It does not provide a certificate with external validity, but it can detect many common problems.
  • Calibration: determines the instrument’s deviation from a traceable reference under controlled conditions and using a documented procedure. The result is a certificate that states the associated uncertainty.
  • Adjustment: physically modifies the instrument to correct a deviation. It is a follow-up action, not an alternative to calibration. An adjusted pipette without calibration before or after the adjustment has not been validated.

ISO 8655 is the reference standard for piston-operated volumetric equipment such as micropipettes, dispensers and digital burettes. It describes the gravimetric test method and the conditions for carrying it out. An ISO/IEC 17025 calibration certificate comes from a calibration laboratory accredited to that standard. This is what auditors require when measurements have regulatory consequences.

When you need a pipette calibration certificate

Set a schedule based on risk

There is no single interval that suits every pipette. Your laboratory should set the frequency according to the risk of an incorrect result. A pipette used every day to prepare dilutions for a quality control product released to a customer does not need the same interval as one used occasionally to dispense buffer in student practicals.

Consider shortening the interval when the pipette is used intensively, by several operators or for many cycles each day. Small volumes also deserve more frequent checks because the relative effect of an error increases as the volume decreases. Take account of aggressive, viscous or volatile liquids and salts that could reach the piston, as well as any explicit documentation requirements in your quality system.

An occasional-use pipette, one used for larger volumes or one used for qualitative applications where volumetric accuracy does not affect the result may need less frequent calibration.

Check the pipette after an incident or change

Verify the pipette before using it again after any of the following:

  • A drop or impact. Check it even if there is no visible damage. The piston shaft may be misaligned.
  • Liquid entering the pipette body. This can happen when pipetting horizontally or releasing the plunger too quickly. Liquid can reach the piston and seal, affecting the delivered volume.
  • Maintenance or replacement of seals, springs or the piston.
  • Autoclaving, if the model is suitable for it. The heat cycle can affect the seals.
  • A change of tip type or supplier. The fit between the cone and tip is part of the measuring system. A poorly sealing tip can introduce error even when the pipette is in good condition.
  • A change of application, such as moving from water to a solvent with a very different density and vapour pressure.
  • Suspicious results, including unusual calibration curve slopes, inconsistent replicates or drift in a control.
  • Before a long study or method validation. Repeat the check at the end to show that the instrument remained within limits during the work.
  • When a new pipette arrives. A factory calibration certificate does not guarantee that the instrument arrived in the same condition as when it left the factory.

Signs that a micropipette may be failing

Some problems are visible at the bench, before you check the pipette on a balance:

  • Drops hang from the tip or fall off when you move the pipette.
  • The tip does not fit securely or falls off. The cone may be worn or dirty.
  • The plunger sticks, has a hard spot or sounds different.
  • The volume setting moves during use.
  • Air bubbles appear repeatedly during aspiration.
  • Two pipettes with the same nominal volume and setting deliver visibly different volumes.

Any of these signs is a reason to take the pipette out of service and check it. None, on its own, means that you need to discard it.

In-house verification or accredited calibration?

In-house verificationAccredited calibration
Who performs itYour laboratoryAn external calibration laboratory
What you needA balance with suitable resolution for the volume, distilled water, a thermometer, barometer, hygrometer and a written procedureSend the instrument to the calibration laboratory
DocumentationAn internal recordA certificate with traceability and stated uncertainty
When it is neededRoutine checks between calibrationsAccredited environments, audits and tests with regulatory consequences

For many laboratories, a combined approach works well: arrange external calibration with a certificate at the interval set by your quality system, then carry out quick in-house checks between calibrations. Prioritise pipettes used every day and those used at low volumes. An external micropipette calibration service can provide the certificate when your quality requirements call for it.

How to carry out a gravimetric verification

The method described in ISO 8655 involves weighing dispensed water and converting mass to volume using a correction factor based on temperature, atmospheric pressure and water density. Avoid these common mistakes:

  • Let the pipette, tips and water reach the laboratory temperature before you start. Temperature differences are a frequent source of error and are easy to prevent.
  • Pre-wet the tip before the first weighing.
  • Work away from draughts and control evaporation, especially at low volumes. A humidity trap or narrow weighing vessel can help.
  • Test several points across the volume range, including the nominal volume, an intermediate point and the minimum. A pipette may meet the limits at the top of its range but fail at the bottom.
  • Repeat the weighings enough times to calculate accuracy, or systematic error, and precision, or spread. One weighing is not enough.
  • Record the operator, environmental conditions, balance and tip batch. Without this information, the record is not useful in an audit.

When you do not need accredited calibration

Not every pipette requires an accredited certificate. Classify pipettes by criticality and focus external calibration on the instruments that need it.

  • You do not need an accredited certificate for every pipette. A pipette used to add an approximate volume of wash buffer or load gels has different accuracy requirements from one used for critical measurements.
  • Do not replace a pipette if the tip is the problem. Poor sealing or drops remaining on the tip can often be addressed by using compatible, good-quality tips. Check the instrument with the correct consumable before deciding it is faulty.
  • Do not replace a pipette for a repairable mechanical problem. Seals, O-rings and springs are wear parts. Maintenance followed by verification may bring the instrument back within specification.
  • Do not buy an electronic pipette just to improve accuracy. Electronic models can improve reproducibility between operators, offer repeat-dispensing programmes and reduce fatigue during long series. For occasional pipetting, a properly calibrated mechanical pipette may be suitable.
  • Do not set up an in-house verification system for only two or three pipettes. The balance, environmental controls and operator time may not be worthwhile; external calibration may be a better option.
  • Do not verify a pipette again just in case immediately after calibration if nothing has happened to it. A quick functional check before a critical test may still be useful.

What to record for a useful calibration history

Validation without a traceable record is work lost. Keep a unique identifier for each pipette, including a physical label, along with its model and volume range. Record the dates of the last and next calibration, accuracy and precision results at each test point, the acceptance criteria, any action taken after a deviation and the person responsible.

Also document what happens when a pipette falls outside the limits. Review which tests have been performed since the last compliant calibration, assess the impact and decide whether work needs to be repeated. This step is often missing and is one auditors ask about.

How to start a pipette calibration plan

If you are setting up a plan from scratch, make an inventory, classify pipettes by criticality, assign different intervals to each group and decide which instruments need an external certificate. If a pipette is already causing problems and is not worth repairing, replace it with one supplied with a calibration certificate and start its record from day one.

Letslab has sold laboratory equipment and supplies online since October 2008, with more than 50,000 catalogue items. See the available models, volume ranges and certificate options in the micropipettes and calibration certificates category. Product pages list the technical specifications and available options.