TAKIVES Semi-Micro ESR System, Automatic Zero
In stock
+150€ Península y Baleares
pago
escuelas y universidades
oficiales
| SKU | Product Name | Availability | Price | Qty | |
|---|---|---|---|---|---|
| 1164 | Citrate tube for ESR 12 x 55 mm, pre-pierced stopper, 1,000 pcs | In stock |
€160.29
|
|
|
| 1361 | Rack for 5 mm tubes, 5 pcs | In stock |
€96.76
|
|
TAKIVES Semi-Micro ESR System with Automatic Zero
Complete single-use system for Erythrocyte Sedimentation Rate (ESR) determination. Complies with Westergren Method recommendations per ICSH reference. Supplied as a complete assembly—citrate tube, automatic zero-lock pipette, and stabilised rack with metal counterweight.
System components
- Tube: High-clarity polypropylene, 11.3 x 56 mm (including stopper)
- Additive: 0.25 ml of 0.109 M trisodium citrate for 1.25 ml total volume (1 ml blood)
- Stopper: Perforable black thermoplastic rubber; max. ext. diameter 15.4 mm; max. tool diameter 6 mm; nominal cut length 7 mm
- Pipette: Integrated piston with automatic zero-lock closure; printed scale for direct reading
- Rack: Plastic, ten-tube capacity with metal counterweight; wide base for vertical stability
- Packaging: Plastic trays, 10x10
Procedure
Fill the tube with blood to the nominal volume and mix with the additive. Complete the determination within the first 4 hours. If the determination is carried out between 4 and 12 hours after collection, store the sample refrigerated at 4 °C and allow to equilibrate to ambient temperature before use.
Prior to determination, mix by at least 12 inversions to ensure homogenisation. Insert the pipette into the tube and raise the piston until blood reaches the automatic zero-lock. Place pipettes vertically in the rack—no vibration, temperature between 18 and 25 °C, away from direct sunlight.
Read at 1 hour; second reading at 2 hours. Results can be read directly from the printed scale on the pipette or via automated methods.
Critical conditions for accurate readings
- Blood volume must correspond exactly to the nominal volume defined for the tube
- Column filling must be continuous—air bubbles will invalidate the result
- Maintain working temperature between 18 and 25 °C throughout the entire determination
Storage and stability
- Temperature: Between 4 and 25 °C (40–77 °F)
- Light: Protect from direct sunlight
- Stability: 15 months from date of manufacture
- Labelling: Pack orientation arrow, identification, lot number and expiry date