Water demineraliser: choosing Type I or Type II lab water

A water demineraliser may be the right choice when your laboratory needs water without salts, but the label distilled water does not tell you what grade you are getting. Distillation describes a process, not a water quality: two distilled waters can differ substantially in conductivity, organic content and microbiological load. For laboratory use, the required grade is what matters, which is why water is commonly described as Type I, Type II or Type III.
What a water demineraliser does, and what it does not
The usual reference for analytical laboratory water is ISO 3696, which classifies water for analytical use into three grades, 1, 2 and 3. It sets limits for conductivity, oxidisable matter, ultraviolet absorbance, residue after evaporation and silica. In everyday laboratory use, these grades are often called Type I, Type II and Type III. Clinical settings also use sector-specific classifications. They are not identical, but follow the same principle: more demanding techniques allow less tolerance for impurities.
Water purity is not defined by a single measurement. Conductivity, or its inverse, resistivity, indicates the level of dissolved ions. Water can have very low conductivity and still cause problems in HPLC if it contains organic compounds that a conductivity meter cannot detect.
Type I water: ultrapure
Type I is the most demanding grade, with virtually no ions and minimal organic content. With the appropriate polishing, it can also be free of particles and nucleases. It is usually produced at the point of use because its quality deteriorates during storage: it absorbs CO₂ from the air, dissolves ions from its container and can support bacterial growth. Type I water is not something to store in a carboy for the next day.
Type II water: analytical grade
Type II is the laboratory workhorse. It is used to prepare reagents and standards, supply routine analytical instruments, feed autoclaves and laboratory dishwashers, and often supply a Type I system. In many laboratories, it replaces the traditional water still.
Type III and general deionised water
Type III water is suitable for washing and rinsing, water baths and equipment top-ups. The goal is to remove salts that leave residue and limescale, not to achieve analytical purity. An ion-exchange deionizer can do this at a much lower installation and running cost.
Which laboratory water grade does each application need?
| Application | Suitable grade | Reason |
|---|---|---|
| Washing and rinsing laboratory glassware | Deionised or Type III | Helps prevent salt residue and limescale |
| Feeding autoclaves, water baths and steam ovens | Deionised or Type II | Helps protect heating elements and circuits from scale |
| Preparing routine solutions and standards | Type II | Supports accuracy in the final concentration |
| UV-Vis spectrophotometry, titrations and general microbiology | Type II | Provides stable blanks without ionic interference |
| HPLC, ion chromatography, ICP-MS and molecular biology | Type I | Organic content and trace ions are critical |
| Feeding an ultrapure water system | Type II | Extends the life of polishing cartridges |
Two approaches: deionising or purifying water
Wall-mounted water deionizer for volume and simplicity
When you need a continuous supply of water without salts, an ion-exchange deionizer is a straightforward option. The Wall-Mounted Water Deioniser 0.1–20 μS/cm 50 l/h from SMEG connects directly to the mains water supply, with no intermediate tank. It removes inorganic substances using ion-exchange resin and delivers up to 50 l/h, with an outlet conductivity range of 0.1 to 20 μS/cm.
Two practical features make it convenient in a laboratory with multiple users. The integrated conductivity meter shows the reading in μS/cm in real time, so you can check whether the water is suitable for its intended use without measuring it separately. The cartridge can be changed without tools, so routine maintenance does not depend on an external technician.
Be clear about what a deionizer does and does not do: it removes ions. It is not designed to guarantee the absence of organic matter or microbiological control. An output of 0.1 μS/cm does not make the unit a Type I water generator.
Type II benchtop water purification system for analytical work
If water will be used in an analytical method, low conductivity alone is not enough. You need a specified water quality. The MICROMATIC Benchtop Water Purification System Type II Analytical Grade 2.5 l/h from Wasserlab produces Type II Analytical Grade water directly from mains water, at a flow rate of 2.5 l/h.
A microprocessor and digital display monitor the final water quality and system status, and provide a consumable replacement alert. The system is factory calibrated under SGC ISO 9001 and includes a cartridge kit to get started. Compared with a traditional benchtop water distiller, it is less costly to run and more efficient in consumption. It also avoids boiler cleaning and the scale build-up that damages heating elements.
Flow rate matters as much as water grade
A common mistake when comparing systems is to look only at purity. A system producing 2.5 l/h and one producing 50 l/h solve different problems.
| DEM-10 (SMEG) | MICROMATIC (Wasserlab) | |
|---|---|---|
| Water produced | Deionised, 0.1 to 20 μS/cm | Type II Analytical Grade |
| Flow rate | Up to 50 l/h | 2.5 l/h |
| Installation | Wall-mounted, direct mains connection, no intermediate tank | Benchtop, supplied from mains water |
| Monitoring | Integrated conductivity meter | Microprocessor, digital display and consumable replacement alert |
| Maintenance | Cartridge changed without tools | No boiler cleaning; system alerts you when consumables need replacing |
| Suitable for | Washing, autoclaves and general water use | Reagents, standards and routine analytical work |
A laboratory that fills wash vessels and feeds an autoclave uses many more litres than one that prepares standard solutions. If daily consumption is high and analytical requirements are modest, a 2.5 l/h purifier may not keep up, even if it produces higher-grade water. Conversely, 50 l/h of deionised water does not replace water of a defined grade when a validated method requires it.
When not to buy a high-grade water system
- If the water goes down the drain. Deionised water is sufficient for rinsing glassware. Using Type II water for washing wastes consumables.
- If no method requires it. A teaching or basic testing laboratory rarely needs an ultrapure water system. The step from Type II to Type I increases running costs.
- If you use only a few litres a week. Buying bottled water of the required quality may be more practical than installing and maintaining an underused system. Cartridges have a limited life even when the system is not producing water.
- If the problem is the feed water. Very hard mains water can exhaust cartridges quickly. Check the pretreatment before moving to a higher-grade system.
Consider the replacement cartridge as well as the system
The system is a one-off purchase; the cartridge is a recurring cost. Before choosing a water demineraliser or other water system equipment for your lab, check which replacement part it uses and how it is fitted. The original Replacement Cartridge for DEM-10 Wall-Mounted Water Deioniser fits into the unit and locks into its working position. It does not require flushing or subsequent adjustment, so the system can resume operation straight away.
The most reliable signal to replace a cartridge is a drop in the quality of the water produced, not the calendar. That is why an integrated conductivity meter is more than a convenience: it can show when a cartridge is exhausted before poor-quality water reaches a sample. If you work with batches and traceability, record the conductivity reading when you change the cartridge and monitor any drift.
Choose a water demineraliser in three questions
- Will the water be used in an analytical result? If yes, specify the grade, with Type II as a minimum, rather than starting with litres.
- How many litres do you actually use each day? Measure consumption for a week before buying. The result is often surprising.
- Where will the system go? A wall-mounted unit frees up bench space and provides flow; a benchtop unit can sit next to the work area where the water is used.
Letslab has sold laboratory equipment online since October 2008 and works with more than 50,000 references. If you are unsure which setup to choose, tell us the analytical technique and estimated water consumption. You can browse available options in the water purification category.
Practical recommendation
If your laboratory prepares reagents, standards and samples for routine analysis and you want to replace a traditional still, consider the MICROMATIC Benchtop Water Purification System Type II Analytical Grade 2.5 l/h from Wasserlab. It produces 2.5 l/h, has microprocessor control with a digital display and a consumable replacement alert, is factory calibrated under SGC ISO 9001, and includes a cartridge kit.
If you need a higher volume of salt-free water for washing, an autoclave and general use, the Wall-Mounted Water Deioniser 0.1–20 μS/cm 50 l/h is the higher-flow option, delivering up to 50 l/h at 0.1 to 20 μS/cm. Use the Replacement Cartridge for DEM-10 Wall-Mounted Water Deioniser for maintenance.